ABSTRACT: Purpose
This study investigated the impact of a higher-than-expected corneal index of refraction (nc) on measured central corneal thickness (CCT) by Scheimpflug tomography and anterior-segment OCT.Design
Theoretical analysis with prospective cohort study validation.Participants
Twenty-four eyes from 23 participants met the criteria for inclusion in data analysis.Methods
Two models were developed to represent side-view imaging with Scheimpflug geometry (model 1) and interferometric imaging in OCT (model 2). For both models, predicted CCT was calculated for various nc values, and CCT measurement error was quantified. For validation, CCT was measured using Scheimpflug tomography and OCT in prospectively recruited subjects with keratoconus