Single-isocenter multiple-target stereotactic radiosurgery (SIMT SRS) has emerged as an efficient treatment for multiple brain metastases. However, this technique demands exceptional geometric accuracy, particularly off-axis, to ensure optimal dose delivery while sparing healthy tissue. Traditional quality assurance (QA) methods require adaptation for SIMT SRS, highlighting the need for robust testing protocols.Purpose
This study aimed to evaluate the long-term off-axis targeting accuracy of three Varian TrueBeam linear accelerators using a dedicated off-axis Winston-Lutz test (OAWLT) and to investigate correlations with routine varian machine performance check (MPC) results.Materials & methods
Weekly OAWLT measurements were performed over a 6-month perio