Influence of geometry, reinforcement, and sterilisation on the dimensional accuracy of additively manufactured carbon fibre-reinforced nylon composites.
Ontology highlight
ABSTRACT: Dimensional accuracy is a critical quality metric in manufacturing, particularly for medical devices subjected to sterilisation and disinfection. While additive manufacturing (AM), especially fused filament fabrication (FFF), facilitates the production of complex geometries, challenges such as void formation, surface deformation, and mechanical instability persist. This study evaluated the impact of sterilisation (autoclaving) and disinfection (ethanol) on the dimensional stability of 3D-printed carbon fibre-reinforced polymer (CFRP) parts. Two geometries - representing standard ASTM D3039 and complex non-standard designs - were printed using carbon fibre nylon-based composites with and without continuous carbon fibre (CCF) reinforcement. Dimensional accuracy and void fraction were assesse
SUBMITTER: Cheers GM
PROVIDER: S-EPMC12489055 | biostudies-literature | 2025 Oct
REPOSITORIES: biostudies-literature
ACCESS DATA