HyperArc Automated Stereotactic Radiosurgery Planning Enables Accurate a Priori Fractionation Scheme Selection via Adherence to HyTEC Toxicity Thresholds.


ABSTRACT:

Purpose

Radiosurgery plan safety is commonly estimated by volumes receiving specific doses (ie, 12 Gy/1 fraction [fx]), which are evaluated postplan generation. However, automated treatment planning can produce highly consistent and thus predictable plans. Thus, we hypothesized that HyperArc (HA) automated stereotactic radiosurgery (SRS) planning enables clinical decision-making prior to plan generation, such as selecting the appropriate SRS fractionation scheme.

Methods and materials

All previously treated single-isocenter HA plans at our institution were queried, totaling 3361 marginless targets without bridging at the 50% isodose level (1495 plans), making this the largest single-institutional SRS dosimetry study to the authors' knowledge. Eight isodose volumes (IDVs; 50.