Unknown

Dataset Information

Digital smartphone versus analogue methods of measuring orbital implant motility after enucleation.


ABSTRACT:

Purpose

To compare standard rule measurement (gold standard, analogue) method of orbital implant motility against digital image analysis on smart phone measurement software, Medigrid and ImageJ.

Design

Single centre observational, non-randomised, non-masked cohort study in enucleated patients.

Methods

A standard rule measurement (gold standard) was used to measure vertical and horizontal movements of the fellow eye, prosthesis, and orbital implant. At the same time, photographs taken in the same positions of gaze were analysed using ImageJ software (NIH, Bethesda, Maryland, USA) and the smartphone app mediGrid (IRISS Medical Technologies, UK). Bland-Altman plots were used to compare the digital modalities, with the current gold standard.

Results

A total of 54 p

SUBMITTER: Gopal BP 

PROVIDER: S-EPMC12494908 | biostudies-literature | 2025 Oct

REPOSITORIES: biostudies-literature