ABSTRACT: Background
Since the introduction of whole-body photon-counting detector CT (PCD-CT) into clinical practice, extensive physics assessments have been conducted to elucidate its image quality advantages over energy-integrating detector CT (EID-CT) and to support its clinical adoption. However, evaluations of its three-dimensional (3D) noise power spectrum (NPS), which simultaneously quantifies in-plane and through-plane noise texture and magnitude, remain limited.Purpose
To experimentally evaluate the 3D NPS of an NAEOTOM Alpha PCD-CT system and its dependence on scan mode, reconstruction image type, quantum iterative reconstruction (QIR) strength, mono-energy (keV) level, spiral pitch, and radiation dose.Methods
Repeated scans of a 20 cm water phantom and a 30 cm PMM