ABSTRACT: Background
This study investigated the use of a respiratory function monitor (RFM) to guide the placement of a supraglottic airway device (SAD) in neonates during intensive care interventions. We hypothesized that using a RFM would decrease the number of attempts needed for a successful placement.Methods
This single-center pilot study was carried out at a tertiary NICU at the Medical University of Vienna. Patients were ventilated using a SAD during neurosurgical or endoscopic interventions. A RFM was either hidden (but recording) or visible to providers during SAD placement. Feedback from the RFM was used to assess correct/incorrect placement and optimize ventilation quality. The parameter leakage was used for assessment: if leak was <30%, correct placement was assumed. The