Acute durability of cavotricuspid isthmus block after pulsed electric field ablation: randomized comparison of two pentaspline catheter configurations (SECTION trial).


ABSTRACT:

Aims

Cavotricuspid isthmus (CTI) ablation is commonly performed alongside catheter ablation of atrial fibrillation (AF). However, the acute efficacy of the CTI ablation using the pentaspline catheter and pulsed electric field (PEF) energy has not been systematically evaluated. This randomized study assessed the acute efficacy and extent of haemolysis associated with CTI ablation when performed using two different configurations of the pentaspline catheter.

Methods and results

A total of 178 patients (age 65 ± 10 years, 66% of males) undergoing PEF ablation of the CTI in conjunction with AF ablation were randomly assigned to receive ablation using either the basket configuration (n = 95) or the flower configuration (n = 83) of the pentaspline catheter. The CTI ablation was pe