ABSTRACT: Background
Terbium-161 (161Tb) is a promising β⁻-emitter for theragnostics. However, its complex photon emission pattern-including intense X-rays and low-yield, high-energy γ-emissions-may complicate image-based quantification. This study aimed to assess the feasibility of accurate SPECT/CT-based 161Tb dosimetry through a series of phantom measurements using a GE Discovery NM/CT 670 Pro system. Three collimators were evaluated: extended low-energy general-purpose (ELEGP), low-energy high-resolution (LEHR), and medium-energy general-purpose (MEGP), using two separate energy windows: around the 75 keV γ-peak (± 10%), and around the 49 keV γ-peak and nearby X-rays (40.7-62.9 keV). A clinical OSEM reconstruction algorithm was employed.Results