Unknown

Dataset Information

Development and tracking of inferior vena cava diameter after filter placement using a circle-fitting predictive model.


ABSTRACT: The diameter of the inferior vena cava (IVC) is a critical factor in the decision-making process for IVC filter (IVCF) placement. A mismatched filter can compromise its stability and potentially lead to adverse events. This study develops and validates a predictive model using a circle-fitting algorithm to estimate the IVC diameter after filter placement. The findings suggest that single-angle measurements may misrepresent the filter-vessel compatibility, since the IVC transforms from an oval to a near-circular cross-section following filter placement. The predictive model, integrating three-dimensional, multi-angle cavography with circumference-based calculations, demonstrates excellent concordance with both post-placement maximum and minimum diameters in animal and clinical samples. Comp

SUBMITTER: Gong M 

PROVIDER: S-EPMC12519178 | biostudies-literature | 2025 Oct

REPOSITORIES: biostudies-literature