Exhaled Carbon Dioxide Monitoring Using Instantaneous CO<sub>2</sub> Mode During High-Frequency Oscillatory Ventilation in Very Low Birth Weight Infants.


ABSTRACT:

Background

High-frequency oscillatory ventilation (HFOV) presents challenges for noninvasive CO2 monitoring due to its rapid oscillations. This study examines the feasibility of instantaneous CO2 partial pressure (INST CO2) monitoring during HFOV with a capnometer equipped with a mainstream CO2 sensor.

Methods

This study included seven neonates receiving HFOV in the neonatal intensive care unit (NICU) at Shizuoka Children's Hospital. INST CO2 values were recorded over 3- and 5-min intervals surrounding each blood gas sampling to determine the maximum INST CO2 value (INST CO2 MX). The primary outcome was the correlation between INST CO2 MX and arterial partial pressure of carbon dioxide (P