Development of a new cell isolation device FlowMagicTM.
Ontology highlight
ABSTRACT: Isolation of human peripheral blood mononuclear cells (PBMCs) from blood typically involves a density gradient medium during density centrifugation. The problem of increasing red blood cell (RBC) and granulocyte (GRA) contamination during PBMC isolation as the elapsed time after blood collection increases remains unresolved. As a countermeasure against RBC contamination, hemolysis treatment is available; however, these extra steps are laborious, time-consuming, and could introduce artifacts. To overcome this challenge, we developed a novel isolation device, FlowMagic™, which features a proprietary two-layer insert structure designed to prevent RBC and GRA contamination during PBMC isolation from blood. The efficacy of this method was evaluated by isolating PBMCs from donors and analyzing i
SUBMITTER: Yoshida T
PROVIDER: S-EPMC12543197 | biostudies-literature | 2025