Predictive value of tidal volume and peak inspiratory pressure in normal frequency jet ventilation.


ABSTRACT: Normal frequency jet ventilation (NFJV) is commonly used during rigid bronchoscopy; however, airway opening does not reliably measure or predict tidal volume (VT) or peak inspiratory pressure (PIP). This study aimed to investigate the comprehensive effects on VT and PIP concerning key variables, including driving pressure (DPpipeline), frequency (Fjet), needle inner diameter (IDneedle), and lung dynamic compliance (Cdyn). Three jet needles (N1, N2, and N3) with different internal diameters (1.2-1.9 mm) were used to deliver jet ventilation via a rigid bronchoscope at DPs of 0.6-1.6 bar and frequencies of 10-60 min⁻¹. Airway pressure (Paw) was measured near the tracheal prominence of the simula