Surface-guided breathing signal integration in breathing-adapted intelligent 4D computed tomography: prototype implementation and comparison with an infrared marker-based system.


ABSTRACT:

Background and purpose

Breathing-adapted intelligent four-dimensional computed tomography (i4DCT) reduces motion artifacts during irregular breathing using real-time surrogate signal analysis to control X-ray timing. Current implementations rely on infrared (IR) markers or pressure belts. Surface-guided radiation therapy (SGRT), a widely used markerless technique in radiotherapy, has not been evaluated for direct i4DCT integration. This study tested SGRT for i4DCT acquisition in a phantom setup and benchmarked it against an infrared marker-based system.

Materials and methods

Phantom measurements with clinically relevant regular and irregular breathing signals were performed on a prototype CT scanner with direct SGRT control. To improve SGRT-based surrogate accuracy, a table