Multilayer and crushed stent visualization using photon- counting detector computed tomography - a proof of concept.


ABSTRACT: Photon‑counting detector CT (PCD‑CT) enables ultra‑high resolution (UHR, 0.2 mm) and may improve visualization of coronary stents compared with standard‑resolution (SR, 0.4 mm). In this ex- vivo study, we utilized a phantom heart model to simulate physiological attenuation and imitated the most common coronary bifurcation stenting techniques by creating multiple stent layers and stent- crush situations (single layer, crushed, two‑ and three‑layer). We compared UHR and SR PCD‑CT using Bv72 and Bv56 kernels. Objective endpoints were in‑stent lumen visibility, signal-to-noise ratio, and percentage change in stent attenuation; subjective image quality was rated on a 5‑point Likert scale (1 = excellent to 5 = non‑diagnostic). Depending on distribution, two‑sided t‑tests (mean ± SD) or Mann-Whit