ABSTRACT: Background
The patency of intracranial stents may not be reliably assessed with either CT angiography or MR angiography due to imaging artifacts. We investigated the potential of ultra-high resolution CT angiography using a photon counting detector (PCD) CT to address this limitation by optimizing scanning and reconstruction parameters.Methods
A phantom with different flow diverters was used to optimize PCD-CT reconstruction parameters, followed by imaging of 14 patients with intracranial stents using PCD-CT. Images were reconstructed using three kernels based on the phantom results (Hv56, Hv64, and Hv72; Hv=head vascular) and one kernel to virtually match the resolution of standard CT angiography (Hv40). Signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) measure