ABSTRACT: Background
The recently developed motion-compensated reconstruction (MCR) algorithm provides a fully automated process to generate motion-corrected images. This study aimed to assess the image quality and radiation dose of prospective electrocardiogram (ECG)-triggered coronary computed tomography angiography (CCTA) using MCR on second-generation dual-layer spectral detector computed tomography (DLCT), and to investigate the influence of heart rate (HR) on the motion-correction effect of this algorithm.Methods
This prospective study enrolled 73 patients who underwent CCTA on DLCT utilizing prospective ECG-gating. Patients were divided into two subgroups with a regular rhythm HR of <75 bpm and HR of ≥75 bpm and/or an irregular heart rhythm, respectively. Both subjective and o