Contrast media and radiation dose optimization with task-based automatic keV selection: a proof-of-concept study with photon-counting detector CT.


ABSTRACT:

Objectives

To evaluate whether task-based automatic keV selection of photon-counting detector (PCD)-CT with optimizing radiation and contrast media (CM) dose yields consistent image quality in CT angiography (CTA).

Materials and methods

PCD-CTA of the aorta was performed in six healthy minipigs across two scan sessions, with virtual monoenergetic images (VMI) reconstructed. In the first session, three protocols were conducted: the reference protocol A1 simulated standard CTA (210 mg iodine/kg CM, image quality (IQ)-level 117, non-contrast task, VMI: 70 keV); protocol A2 reduced radiation while keeping CM dose constant (210 mgI/kg, IQ-level 117, vascular task, VMI: 55 keV); and protocol A3 reduced CM dose while maintaining radiation (164 mgI/kg, IQ-level 117, non-contrast tas