Arterial-optimized 4D-flow MRI for quantifying flow and pulsatility in venous sinuses and large cerebral veins.
Ontology highlight
ABSTRACT: A single, arterial-optimized 4D-flow MRI acquisition may enable fast assessment of both cerebral arterial and venous flow. However, arteries and veins require different velocity encoding (VENC) settings for optimal velocity-to-noise ratio (VNR). Consequently, venous measurements using arterial-optimized VENC settings are subject to reduced VNR and require further evaluation. This study compared cerebral venous flow and pulsatility assessments using a high-VENC (110 cm/s, adapted to the arterial system) and a low-VENC (40 cm/s, adapted to the venous system) 4D-flow MRI sequence at 3 Tesla. Flow and pulsatility index (PI) were calculated for cerebral veins, sinuses and internal jugular veins in 36 elderly volunteers (79 ± 5 years). The high-VENC acquisitions allowed visualization of nearly a