Ontology highlight
ABSTRACT: Purpose
This study prospectively compared the diagnostic accuracy of vibration-controlled transient elastography (VCTE) and point shear wave elastography (pSWE) in the assessment of liver fibrosis, versus a reference standard of magnetic resonance elastography (MRE).Methods
This study prospectively enrolled patients with chronic liver disease, who underwent pSWE, VCTE, and MRE. Fibrosis was staged based on liver stiffness (LS) values measured using MRE: F0 (<2.61 kPa), F1 (≥2.61 to <2.97 kPa), F2 (≥2.97 to <3.62 kPa), F3 (≥3.62 to <4.69 kPa), and F4 (≥4.69 kPa). Each modality was performed independently, and the results were blinded to minimize bias. Diagnostic performance was assessed using the Pearson correlation coefficient (CC), Lin concordance correlation coefficient (
SUBMITTER: Gotoh T
PROVIDER: S-EPMC12645073 | biostudies-literature | 2025 Nov
REPOSITORIES: biostudies-literature