Comparison of Distortion-Product Otoacoustic Emissions Measured in the Same Subjects Using Four Commercial Systems.
Ontology highlight
ABSTRACT: Background/Objectives: Distortion-product otoacoustic emissions (DPOAEs) are suited to longitudinal cochlear assessment, but inter-system differences may confound interpretation across clinics. This study compared DPOAE outcomes across four commercial systems and evaluated their within-session repeatability. Methods: Adults with normal hearing (84 ears) were tested using the HearID DP (Mimosa Acoustics), SmartDPOAE (Intelligent Hearing Systems), Eclipse DPOAE20 (Interacoustics), and Echoport ILO 292 USB I (Otodynamics). DPOAEs were recorded at 1, 1.5, 2, 3, 4, 6 and 8 kHz using a criterion of ≥6 dB signal-to-noise ratio. Two measurements per ear were obtained, with the probe repositioned between sessions. Results: All systems showed similar frequency response profiles