Automated Imaging and Analysis of Platelet, Coagulation and Fibrinolysis Activities Using a Novel Flow Chip-Based System at Physiological Temperature.
Ontology highlight
ABSTRACT: Conventional whole-blood flow assays for quantifying thrombus formation are typically performed at room temperature and are technically demanding, which limits their translational applicability. We engineered a novel, disposable, mountable, and single-channel microfluidic chip (MC-2S), which is based on the Maastricht chamber (MC) and designed for automated evaluation of platelet function, coagulation and fibrinolysis under physiological conditions. The MC-2S chip allows customizable choices of thrombogenic surfaces, such as collagen and tissue factor. The chip was used in combination with an adapted, 1.3 kg brightfield/fluorescence microscope, operating at physiological temperature (37 °C), and with scripts for automated multicolor analysis of image features. The integrated system enables
SUBMITTER: Gui X
PROVIDER: S-EPMC12654719 | biostudies-literature | 2025 Oct
REPOSITORIES: biostudies-literature
ACCESS DATA