Diagnostic Performance of a Silver-Amplified vs. a Non-Amplified Lateral Flow Kit for Adenoviral Conjunctivitis: A Multicenter Prospective Study.


ABSTRACT: Accurate diagnosis of adenoviral conjunctivitis is critical for timely treatment and infection control. However, conventional lateral flow kits often lack sufficient sensitivity, especially in mild cases. This multicenter prospective study evaluated the diagnostic performance of silver-amplified lateral flow kits (SA-LFKs) compared with non-silver-amplified kits (NSA-LFKs), using real-time polymerase chain reaction (qPCR) as the reference standard. Tear samples from 200 patients with suspected adenoviral conjunctivitis were collected across four clinics and analyzed for sensitivity, specificity, and genotype coverage. The SA-LFK demonstrated significantly higher sensitivity (86.0%) than the NSA-LFK (72.0%) (p < 0.001), with both kits showing high specificity (100% and 98.1%, respect