Unknown

Dataset Information

Impact of the maximum ring difference on image quality and noise characteristics of a total-body PET/CT scanner.


ABSTRACT: The sensitivity of a PET system highly depends on the axial acceptance angle or maximum ring difference (MRD), which can be particularly high for total-body scanners due to their larger axial field of views (aFOVs). This study aims to evaluate the impact on image quality (IQ) and noise performance when MRD85 (18°), the current standard for clinical use, is increased to MRD322 (52°) for the Biograph Vision Quadra (Siemens Healthineers).

Methods

Studies with a cylindrical phantom covering the 106 cm aFOV and an IEC phantom filled with 18F, 68Ga and 89Zr were performed for acquisition times from 60 to 1800 s and activity concentrations from 0.4 to 3 kBq/ml to assess uniformity, contrast recovery coefficients (CRCs) and to characterize noise by coefficie

SUBMITTER: Schmidt FP 

PROVIDER: S-EPMC12664500 | biostudies-literature | 2025 Aug

REPOSITORIES: biostudies-literature