ABSTRACT: Purpose
To perform experimental as well as independent Monte Carlo (MC) evaluation of the MC algorithm implemented in RADIANCE version 4.0.8, a dedicated treatment planning system (TPS) for 3D electron dose calculations in intraoperative radiation therapy (IOERT).Methods and materials
The MOBETRON 2000 (IntraOp Medical Corporation, Sunnyvale, CA) IOERT accelerator was employed. PDD and profiles for five cylindrical plastic applicators with 50-90 mm diameter and 0°, 30° beveling were measured in a water phantom, at nominal energies of 6, 9 and 12 MeV. Additional PDD measurements were performed for all the energies without applicator. MC modeling of the MOBETRON was performed with the user code BEAMnrc and egs_chamber of the MC simulation toolkit EGSnrc. The generated phase s