Real-World Safety Data From the World Apheresis Association Registry for the Spectra Optia Apheresis System.
Ontology highlight
ABSTRACT: In this study we analyzed 12 years of adverse events, (AE) data from the World Apheresis Association, (WAA) Registry specific to the Spectra Optia Apheresis System. We queried WAA Registry data, (2012-2023) on apheresis procedures performed exclusively on the Spectra Optia Apheresis System. We categorized AEs by severity, (mild, moderate, and severe), and ordered them by year. We then analyzed and presented AEs associated with the following variables: diagnosis, procedure type, vascular access, and replacement fluid type, and causes of procedural interruption. We identified 51 567 apheresis procedures specific to the Spectra Optia Apheresis System within the data set. AE rates from 2012 to 2023 for mild, moderate, and severe were 1.43%, 2.81%, and 0.21%, respectively. Procedures associated
SUBMITTER: Haubner A
PROVIDER: S-EPMC12665877 | biostudies-literature | 2025 Dec
REPOSITORIES: biostudies-literature