Ontology highlight
ABSTRACT: Purpose
Using parallel transmission (pTx) in ultra-high field MRI allows efficient radiofrequency pulse designs that can produce uniform excitations. Here, we demonstrate that pTx can simultaneously mitigate B1+$$ {\mathrm{B}}_1^{+} $$ non-uniformity and off-resonance artifacts in balanced steady-state free precession sequences at 7 T.Theory and methods
The standard hard excitation pulses of the balanced steady-state free precession sequence were replaced with two pTx pulses, played out in alternating repetition times. Based on the kT-point trajectory, each pulse was designed to produce spatially varying phase distributions that compensate for off-resonance phase variations accumulated over one repetition time duration, while also mitigating B1+$$ {\mathrm{B}}_1^
SUBMITTER: Wu CY
PROVIDER: S-EPMC12681311 | biostudies-literature | 2026 Feb
REPOSITORIES: biostudies-literature