ABSTRACT: Purpose
Transcatheter aortic valve replacement (TAVR) is an established treatment for severe aortic stenosis; however, it carries the risk of periprocedural strokes. Current cerebral embolic protection (CEP) devices, such as the Sentinel, provide partial protection but are limited by inadequate anatomical coverage and inability to capture smaller emboli effectively. This study aimed to evaluate the effectiveness of a novel CEP device, the F2 filter with a 28 μm pore size and full cervical vessel coverage, in preventing emboli from entering the cerebral circulation.Methods
The Sentinel and F2 filter were evaluated for the ability to prevent embolic particles of various sizes (45-300 µm) from entering cerebral arteries using two in vitro flow models, incorporating standard an