Assessing the efficacy of a low-cost air pollution monitoring device for environmental and occupational exposure assessments.
Ontology highlight
ABSTRACT: Low-cost particulate matter (PM) sensors are increasingly used for personal and environmental air quality monitoring due to their affordability and accessibility. Recent advancements make these sensors suitable for occupational settings, but their accuracy in such settings remains uncertain. This study calibrated the AirBeam 2 and AirBeam 3 against the Thermo Scientific Personal DataRAM PDR-1500 to assess their efficacy at measuring high PM concentrations, such as those in occupational exposure settings, using engine exhaust and biomass smoke as PM sources. Laboratory calibrations were conducted using a sealed chamber. Linear and polynomial regressions assessed agreement with the PDR-1500, while breakpoint analyses identified thresholds where sensor performance shifted. Field calibrations