Ontology highlight
ABSTRACT: Purpose
To evaluate a new device, the Goldmann applanation tonometer (GAT) mounted on a novel handheld 3D-printed modular slit lamp (MSL), and compare it with the gold standard slit-lamp-mounted GAT and the handheld Perkins tonometer.Methods
One hundred eyes from 100 participants were included in this cross-sectional observational study. The MSL mount was customized for the selected tonometer (Model AT030, Carl Zeiss, India). A Samsung Galaxy S2 mobile phone was used for observing mires and photography during tonometry. One observer measured intraocular pressure (IOP) with slit-lamp GAT, and another observer measured with MSL GAT and Perkins tonometers. The tonometry sequence was slit-lamp GAT, MSL GAT, and Perkins tonometer, with a 5-minute interval between readings.Re
SUBMITTER: Anche K
PROVIDER: S-EPMC12707405 | biostudies-literature | 2025 Dec
REPOSITORIES: biostudies-literature