ABSTRACT: Background
HyperSight CBCT offers rapid image acquisition and enhanced image quality, supporting its integration into online adaptive radiotherapy (OART). However, its performance under respiratory-induced tumor motion requires further evaluation.Purpose
This study evaluates the feasibility of using rapid-acquisition HyperSight CBCT for internal target volume (ITV) delineation and dose calculation in the setting of respiratory motion, with implications for OART workflows.Methods
A dynamic thoracic phantom simulating superior-inferior tumor motion (5-25 mm amplitude from the motion center, 6-second cycle time) was imaged using Varian's HyperSight CBCT on the Ethos platform. Four reconstruction algorithms were evaluated: filtered back projection (FDK), iterative CBCT