Real-world glycemic outcomes of a tubeless automated insulin delivery system: a single-center observational study in Italy.


ABSTRACT:

Background

The objective of this study is to describe the short-term change observed in CGM-related measures and relevant clinical variables in individuals with type 1 diabetes transitioning to Omnipod 5 insulin treatment within a real-world setting.

Methods

The study involved adults with type 1 diabetes treated with Omnipod 5, whose data were collected over a 14-days observation period prior to (Time 0) and following the (Time 1) initiation of the patch pump use.

Results

A total of 20 adults with well-controlled glycemia were included in the study. From baseline to follow-up, Time in Range (TIR) significantly increased from 57.3% to 67.3% (P<0.001). Concurrently, there were significant decreases in Time Above Range (TAR) Level 1 (mean difference, -4.7 ± 6.1%, P = 0.0

SUBMITTER: Angelino S 

PROVIDER: S-EPMC12745233 | biostudies-literature | 2025