ABSTRACT: Purpose
The aim was to improve the sensitivity and robustness against B0 inhomogeneities of deuterium metabolic imaging (DMI) using phase-cycled balanced SSFP (bSSFP) methods at 9.4 T.Methods
We investigated two variants of phase-cycled bSSFP acquisitions, namely uniformly weighted multi-echo and acquisition-weighted chemical shift imaging (CSI) to improve the SNR of DMI in the brain after oral [6,6'-2H2]-glucose intake. Phase-cycling was introduced to reduce the off-resonance sensitivity of bSSFP, incurring a moderate SNR loss. Two SNR optimal methods for obtaining metabolite amplitudes from the phase-cycled bSSFP data were proposed. The SNR performance of the two bSSFP variants was compared with the SNR-optimized vendor's standard CSI. Add