Ontology highlight
ABSTRACT: Objective
To assess the performance of a novel microfluidic sperm selection device, SwimCount Harvester (SCH), compared with the Swim-up (SU) method, analyzing both sperm quality and the laboratory's key performance indicators.Design
Prospective, double-blind study using sibling oocytes.Subjects
This study included 100 patients undergoing intracytoplasmic sperm injection. Both partners had normal karyotypes. Male participants provided only fresh ejaculated semen samples without total asthenozoospermia, globozoospermia, azoospermia, or cryptozoospermia. Female partners had at least 2 mature oocytes retrieved.Exposure
Semen samples were analyzed to assess concentration, total progressive motile count, vitality, morphology, chromatin structure integrity, and de
SUBMITTER: Meseguer F
PROVIDER: S-EPMC12746890 | biostudies-literature | 2025 Dec
REPOSITORIES: biostudies-literature