Optimising electrical vestibular stimulation (EVS) for assessing vestibular function.


ABSTRACT:

Objective

Electrical vestibular stimulation (EVS) exhibits similar diagnostic performance to caloric irrigation for assessing interaural vestibular asymmetry. To optimise EVS for clinical uses, we investigated the influence of EVS stimulus parameters upon absolute torsional eye response amplitude and asymmetry assessments.

Methods

In young healthy adults (n = 72), monoaural sinusoidal EVS was delivered to each mastoid process for 40 s whilst ocular torsion was assessed. Ambient lights, electrode montage, and stimulus frequency were systematically manipulated.

Results

Ambient lights (250 lx) attenuated torsional responses by 20 % compared to darkness. When the reference electrode was next to C7 (i.e., mastoid-C7 montage), responses to monoaural stimuli were 50 % those