ABSTRACT: Rationale
The Delta Ray Isotope Ratio Infrared Spectrometer enables to determine δ13C and δ18O values of CO2. The measurements are calibrated against internal reference gas standards. For discrete carbonate or dissolved inorganic carbon (DIC) samples, the manufacturer recommends a calibration against two isotopically different carbonate standard materials (i.e., a 2-point calibration). Here, we show that this method is not sufficient because the measured values may drift significantly over the duration of a measurement run (i.e., 58 samples in approximately 24 h).Methods
We developed a new measurement routine applying a 3-point calibration and a tight standard-bracketing. In addition, we present an R code to automatically evaluate and calibr