Shorter screw lengths in dynamic Dynesys fixation have less screw loosening: From clinical investigation to finite-element analysis.


ABSTRACT:

Background

The dynamic Dynesys Stabilization System preserves lumbar mobility at instrumented levels. This study investigated the effect of screw length on screw loosening (SL) after dynamic Dynesys fixation and screw displacement during lumbar motion, using clinical investigation and finite-element (FE) analysis.

Methods

Clinical data of 50 patients with degenerative spondylolisthesis treated with decompression and Dynesys fixation in 2011 were analyzed retrospectively. Horizontal sliding displacement and vertical displacement of screw tips at L4 were analyzed postoperatively using displacement-controlled FE analysis at the L4-L5 level with screw lengths 45 (long screw), 36 (median screw), and 27 (short screw), and 6.4 mm in diameter, under flexion, extension, lateral bendi