Unknown

Dataset Information

Comparison of Handheld Retinal Imaging with ETDRS 7-Standard Field Photography for Diabetic Retinopathy and Diabetic Macular Edema.


ABSTRACT:

Purpose

To compare nonmydriatic (NM) and mydriatic (MD) handheld retinal imaging with standard ETDRS 7-field color fundus photography (ETDRS photographs) for the assessment of diabetic retinopathy (DR) and diabetic macular edema (DME).

Design

Prospective, comparative, instrument validation study.

Subjects

A total of 225 eyes from 116 patients with diabetes mellitus.

Methods

Following a standardized protocol, NM and MD images were acquired using handheld retinal cameras (NM images: Aurora, Smartscope, and RetinaVue-700; MD images: Aurora, Smartscope, RetinaVue-700, and iNview) and dilated ETDRS photographs. Grading was performed at a centralized reading center using the International Clinical Classification for DR and DME. Kappa statistics (simple [K], weighted

SUBMITTER: Salongcay RP 

PROVIDER: S-EPMC12764358 | biostudies-literature | 2022 Jul

REPOSITORIES: biostudies-literature