ABSTRACT: Background
Battery longevity is a key consideration in the selection of implantable cardioverter-defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds), with implications for both clinical outcomes and healthcare costs. A recent study suggested shorter-than-expected battery life for BIOTRONIK devices, prompting a comprehensive internal analysis. This analysis aimed to evaluate the real-world battery longevity of BIOTRONIK high-voltage devices (ICDs and CRT-Ds) across three generations, using global remote monitoring data.Methods
A total of 215,471 BIOTRONIK devices implanted between 2006 and 2024 were analyzed using data from the BIOTRONIK Home Monitoring Service Center. Devices were grouped into three generations (BIO 2000, 2010, 2020) and stra
SUBMITTER: Hayes D
PROVIDER: S-EPMC12764575 | biostudies-literature | 2026 Jan