ABSTRACT: Objectives
To evaluate the stability of radiomic features under different CT acquisition settings and investigate its impact on diagnostic model performance and generalizability.Materials and methods
198 patients with 1227 pulmonary nodules underwent chest CT scans using varied settings (three slice thicknesses, two reconstruction matrices, six convolution kernels, two transmission methods). 1394 radiomic features were extracted per nodule. Feature stability was evaluated using the Intraclass Correlation Coefficient (ICC, stable: ICC ≥ 0.8, intermediate stable: 0.4 < ICC < 0.8, unstable: ICC ≤ 0.4). Four diagnostic models (Full-feature, Stable, Unstable, Intermediate stable) were developed using two datasets (lung cancer screening, n = 184; clinical scenarios, n = 1192). In