A 3D printed capillary holder design for reliable stopped-flow SAX experiments.
Ontology highlight
ABSTRACT: In this paper we discuss adaptations to the BioLogic stopped-flow module (SFM)-400 stop-flow rapid mixing apparatus to enhance reliability for use on a small angle X-ray scattering (SAXS) beamline. Issues with the standard capillary holders are discussed and a 3D printed alternative is presented. The reliability of the new design against leakage is reported. Alternate resins with enhanced solvent tolerance are trialled and discussed.
SUBMITTER: Wady P
PROVIDER: S-EPMC12809459 | biostudies-literature | 2026 Jan
REPOSITORIES: biostudies-literature
ACCESS DATA