Combined Multiharmonic Frequency Analysis for Improved Dynamic Energy Measurements and Accuracy in Charge Detection Mass Spectrometry.


ABSTRACT: The ability to determine ion energies in electrostatic ion-trap-based charge detection mass spectrometry (CDMS) experiments is important for the accurate measurement of individual ion m/z, charge, and mass. Dynamic energy measurements throughout the time an ion is trapped take advantage of the relationship between ion energy and the harmonic amplitude ratio (HAR) composed from the fundamental and second harmonic amplitudes in the Fourier transform of the ion signal. This method eliminates the need for energy-filtering optics in CDMS and makes it possible to measure energy lost in collisions and changes in ion masses due to dissociation. However, the accuracy of the energy measurement depends on the signal-to-noise ratio (S/N) of the amplitudes used to determine the HAR. Here,