Unknown

Dataset Information

Assessing the Impact of Imaging Parameters on MRI Measurement of Kidney T<sub>2</sub>.


ABSTRACT:

Background

T2-mapping has promise to evaluate kidney pathophysiology. Prior studies show a large variance in kidney T2, likely due to the differing acquisition sequences.

Purpose

To compare four T2-mapping sequences to investigate kidney T2.

Study type

Phantom and prospective in vivo assessments.

Participants

ISMRM/NIST and QASPER phantoms; 8 healthy volunteers (4 female, 30 ± 8 years).

Field strength/sequence

3 T, spin echo-echo planar imaging (SE-EPI), multi-echo-spin echo (ME-SE), Gradient and Spin Echo (GraSE) vendor-provided sequences, and custom T2-prepared EPI, Dual-echo B0-mapping and DREAM B1-mapping.

Assessment

T2-mapping accuracy in the ISMRM/NIST ph

SUBMITTER: Daniel AJ 

PROVIDER: S-EPMC12810998 | biostudies-literature | 2026 Feb

REPOSITORIES: biostudies-literature