ABSTRACT: Background
T2-mapping has promise to evaluate kidney pathophysiology. Prior studies show a large variance in kidney T2, likely due to the differing acquisition sequences.Purpose
To compare four T2-mapping sequences to investigate kidney T2.Study type
Phantom and prospective in vivo assessments.Participants
ISMRM/NIST and QASPER phantoms; 8 healthy volunteers (4 female, 30 ± 8 years).Field strength/sequence
3 T, spin echo-echo planar imaging (SE-EPI), multi-echo-spin echo (ME-SE), Gradient and Spin Echo (GraSE) vendor-provided sequences, and custom T2-prepared EPI, Dual-echo B0-mapping and DREAM B1-mapping.Assessment
T2-mapping accuracy in the ISMRM/NIST ph