Unknown

Dataset Information

Assessing Quality and Adherence to PI-RADSv2.1 Minimum Technical Standards of Prostate MRI in NRG-GU005.


ABSTRACT:

Background

Multi-parametric MRI (mpMRI) datasets often vary between sites due to differences in acquisition protocols.

Purpose

Evaluate adherence of multi-site mpMRI dataset to minimum technical standards (MTS) of PI-RADSv2.1.

Study type

Prospective.

Subjects

Six hundred patients (Age (years): ≤ 49 = 0.8%, 50-59 = 10.7%, 60-69 = 47.0%, ≥ 70 = 41.5%) with intermediate-risk prostate cancer (PCa) imaged across 124 institutions prior to radiotherapy.

Field strength/sequence

3T, 1.5T, and 1.16T, T2-weighted (T2w): fast spin-echo, diffusion-weighted imaging (DWI): single-shot echo-planar imaging, and dynamic contrast-enhanced (DCE): T1-weighted 3D fast spoiled gradient echo.

Assessment

Scanner vendors included Siemens, GE, Philips, Toshiba, and Hitachi.

SUBMITTER: Alley S 

PROVIDER: S-EPMC12811001 | biostudies-literature | 2026 Feb