Technological and Biological Reliability, and Validity of Five Different CPET Systems During Simulated and Human Exercise.
Ontology highlight
ABSTRACT: The validity and between-day reliability of cardiopulmonary exercise testing (CPET) systems remain largely unexplored. We therefore evaluate the validity and between-day technological and biological reliability of five popular CPET systems for assessing respiratory variables, substrate use, and energy expenditure during simulated and real human exercise. The following systems were assessed: Vyntus CPX, Oxycon Pro, VO2 Master, KORR, and Calibre. A metabolic simulator was used to simulate breath-by-breath gas exchange. The values measured by each system (minute ventilation (V̇E), breathing frequency (BF), oxygen uptake (V̇O2), carbon dioxide production (V̇CO2), respiratory exchange ratio (RER), energy from carbohydrates and fats, and total energy expenditure) were compa