ABSTRACT: Purpose
The aim is to evaluate the effectiveness of timed breathing in reducing respiratory motion artifacts in renal chemical exchange saturation transfer (CEST) MRI and to assess potential differences in CEST effects between renal compartments.Methods
An electro-pneumatic phantom with a kidney CEST model simulated variable respiratory motion and sequence-synchronized breathing. Motion-induced deviations from a static reference were quantified using the mean absolute error (MAE). Ten healthy volunteers (six females, four males; 25.2 ± 1.9 years) and one patient (47 years) with ccRCC (3.0 × 2.2 × 2.2) cm3 were examined on a 3 T MRI system using a multi-echo gradient echo sequence with 15 Gaussian-shaped saturation pulses (B1 = 1.5 μT, tp = t