Unknown

Dataset Information

Wireless Resonators With Coupled Versus Decoupled Units: Which Enhances Local SNR of RF Receive Arrays Better?


ABSTRACT:

Purpose

To compare two identically sized wireless resonator designs, one with strongly coupled units and the other with decoupled units, for their ability to enhance receive performance in MRI when used with local receive arrays.

Methods

Both wireless resonator designs were fabricated and experimentally evaluated for detuning efficiency, SNR improvements, and parallel imaging performance (g-factor) at 1.5 T. They were used alongside a 12-channel head receive array, with the standard body coil serving as the RF transmitter.

Results

Experimental data showed that the wireless resonator with decoupled units consistently outperformed that with coupled units, with up to threefold improvement in SNR and a reduction of maximum/average g-factor from 4.6/1.8 to 3.1/1.3. Notably

SUBMITTER: Lu M 

PROVIDER: S-EPMC12850623 | biostudies-literature | 2026 Apr

REPOSITORIES: biostudies-literature