ABSTRACT: Purpose
This study sought to determine the intrasession repeatability of the diffusion-weighted (DW) arterial spin labeling (ASL) sequence at different postlabel delays (PLDs).Methods
We first performed numerical simulations to study the accuracy of the two-compartment water exchange rate (Kw) fitting model with added Gaussian noise for DW PLDs at 1500, 1800, and 2100 ms. Ten young, healthy participants then underwent a structural T1 scan and two intrasession in vivo DW ASL scans at each PLD on a 3T MRI. The Kw, arterial transit time (ATT), and cerebral blood flow maps were linearly registered to the structural images, which were then segmented using FreeSurfer into masks with 35 bilateral gray-matter regions.Results
Simulation results showed that the Kw fitting model performed at an error rate less than 10% at physiological ATTs and Kw values, but that error and bias increased at a PLD of 2100 ms and at ATT ranges where the overall blood signal fraction (A1) is low. In vivo analysis showed a significant positive correlation between intrasession measurements of regional Kw at a DW PLD of 1800 ms (β = 0.33, p < 0.001) only. Furthermore, a significant positive relationship between Kw and cerebral blood flow was seen at a DW PLD of 1500 ms (β = 0.26, p = 0.005) and DW PLD of 2100 ms (β = 0.39, p = 0.006).Conclusion
Overall, DW ASL provides the strongest intrasession repeatability at a PLD of 1800 ms in young, healthy subjects, and a simulation study shows accurate Kw fits at physiologic range of ATTs and Kw values.
SUBMITTER: Zhu YD
PROVIDER: S-EPMC12938883 | biostudies-literature | 2026 Feb
REPOSITORIES: biostudies-literature

Magnetic resonance in medicine 20250910 2
<h4>Purpose</h4>This study sought to determine the intrasession repeatability of the diffusion-weighted (DW) arterial spin labeling (ASL) sequence at different postlabel delays (PLDs).<h4>Methods</h4>We first performed numerical simulations to study the accuracy of the two-compartment water exchange rate (Kw) fitting model with added Gaussian noise for DW PLDs at 1500, 1800, and 2100 ms. Ten young, healthy participants then underwent a structural T<sub>1</sub> scan and two intrasession in vivo D ...[more]