Implementation of an Intraoperative Augmented Reality Environment for Custom-Made Partial Pelvis Replacements-A Proof of Concept and Initial Results.


ABSTRACT: Background: The use of augmented reality (AR) in orthopaedics is growing rapidly but is mainly limited to pre-operative planning and teaching. This study is one of the first to describe the intraoperative application within revision arthroplasty for the positioning of customised partial pelvic replacements. Methods: In a proof-of-concept study an AR environment was used during surgery in 11 cases to enhance implant positioning. Postoperatively, a voxel-based CT deviation analysis was carried out to determine the COR deviation and the cup plane deviation angle. Additionally, digital implant superimposition was conducted. Results: Implantation was possible in all cases with a mean COR deviation vector of 4.2 (SD 2.5; 1.2-9.3) mm and a cup plane deviation angle of 4.4 (SD