Mathematical Model analysis of Wallstent and Aneurx: dynamic responses of bare-metal endoprosthesis compared with those of stent-graft.
Ontology highlight
ABSTRACT: We performed this study in order to analyze the mechanical properties of bare-metal Wallstent endoprostheses and of AneuRx stent-grafts and to compare their responses to hemodynamic forces. Mathematical modeling, numerical simulations, and experimental measurements were used to study the 2 structurally different types of endoprostheses. Our findings revealed that a single bare-metal Wallstent endoprosthesis is 10 times more flexible (elastic) than is the wall of the aneurysmal abdominal aorta. Graphs showing the changes in the diameter and length of the stent when exposed to a range of internal and external pressures were obtained. If the aorta is axially stiff and resists length change, a force as large as 1 kg can act in the axial direction on the aortic wall. If the stent is not firmly
SUBMITTER: Canic S